2 using the menu system, Table 3-18. menu functions, 1 speaker on/off – KEPCO MBT Series User Manual
Page 81: 2 controls lock-out, Using the menu system -43, Speaker on/off -43, Controls lock-out -43, 18 menu functions -43, R.s 3.2.2 and 3.2.2.2, R.s 3.2.2 a

MBTSVC111609
3-43
3.2.2
USING THE MENU SYSTEM
To enter the Menu, press the MENU ON/OFF Key. The functions listed in Table 3-18 are avail-
able and can be displayed one at a time by pressing NEXT, or SHIFT then NEXT (PREVIOUS),
to “navigate” through the Menu options. To select the displayed option, press ENTER key.
3.2.2.1
SPEAKER ON/OFF
1.
Press MENU ON/OFF to enter the Menu, then press NEXT until “Speaker On/Off” is
displayed:
press ENTER to select this function.
2.
Screen displays either “Speaker Enabled” or “Speaker Disabled”. Use the slew keys
to toggle between the two options.
3.
Once the unit is set to the desired option, press MENU ON/OFF key to return to menu
level; press it again to exit the menu system.
3.2.2.2
CONTROLS LOCK-OUT
This option is used to lock-up (disable) the VOLTAGE and CURRENT controls and the slew
keys. When locked, the voltage and current parameters can only be modified by numeric key-
pad entry using Vadj and Iadj keys.
1.
Press MENU ON/OFF to enter the Menu, then press NEXT until “Control lockout”
message is displayed; press ENTER to select this function.
TABLE 3-18. MENU FUNCTIONS
DISPLAY
FUNCTION
Speaker On/Off
Enables or disables audible beep heard when keypad keys are pressed (factory default =
speaker on).
Control lockout
Disables VOLTAGE and CURRENT controls (factory default = controls unlocked).
Remote channel
Enter Remote mode to control another power supply from keypad.
Calibration
Provides procedure for calibrating the power supply.
OVP, OCP setup
Establishes Overvoltage and Overcurrent protection values.
Slew key set-up
Establishes the increment for voltage/current increase/decrease produced by one press of a
slew key.
Rem Dsp Blanking
Disables display updateing during remote operation
GPIB addressing
Allows selection of primary or secondary GPIB addressing
Compatible Mode
Allows selection of Compatibility mode for certain features
Command Language
Allows selection of SCPI or CIIL for remote programming
Speaker On/Off
Next/Enter