343
1
0
Supervisor Setup
Calibrating the display of magnification value
The video window image magnification
is displayed in the bottom-right corner
of the QC5000 main screen.
The image magnification depends on the
optical characteristics of the system and
must be found by performing the mea-
surement procedure provided below.
Calibration procedure
Step 1
Verify that the X pixel size... field is set to 1.000
Step 2
Click Tools, Customize and then click
the Status bar tab. True magnification
must be enabled. If the True magnifica-
tion box is not checked, click on True
magnification and then click Show.
Step 3
Click OK to close the Customize screen, and then return to the Tools/Options/VED tab.
Video Probe Setup