Changing probe size
Click and drag an edge of any probe except the Simple and Worm probes to change the probe size using
the closed fist mouse cursor. The closed fist icon becomes available when the mouse cursor is positioned
over a probe edge.
Click and drag an end of the Simple (or Worm) probe to change probe
size using the pointer finger icon discussed earlier.
Changing scan direction
Scan direction becomes important when the image contains more noise on one side of the edge being
measured than on the other. The scan direction should always be from the cleanest side of the edge to the
noisiest. Scan direction is indicated by the arrow(s) associated with each probe.
Click and drag an end or edge of a video probe using the pointer finger or closed fist icon to change the
scan direction of the probe.
Click and drag a probe edge using the
closed fist icon...
to change the probe size
Click and drag the inner and outer circle probe edges ...
to change the probe size
Click and drag the probe ends or probe edges to change scan direction