Setting up the video probes
Many video measurement and display parameters
are configured using tools provided in the VED
screen.
Click the Tools/Options/VED menu item to display
the VED screen.
Aligning the probe automatically and the high accuracy mode
VED probe alignment can affect the quality of feature data. Properly aligned probes optimizes feature
measurement results.
Check the Realign probes... box to automatically optimize
the alignment of VED probes when they’re fired.
Specifying the advanced teach mode
Video probe edge recognition is optimized using either the Standard or Advanced edge teach tool.
The Standard edge teach tool automatically op-
timizes video edge recognition based on 1 of 4
specified edge algorithms; First, Strongest, Auto
or Soft.
Poorly aligned probe
Realigned probe