5-6 DynaRes - Technical Notes
11-13-01
DataShuttle and DynaRes
Product Specifications: DynaRes 8/16
Specifications of the DynaRes 8 and DynaRes 16
Accuracy DynaRes 8/16 General Conditions
Range
Typical Resolution
at Full Scale
Absolute Accuracy, the Larger of
% of Range: % of
Reading:
-5 to +50 mV
0.8 µV
0.04%
–
-25 to +25mV
0.8 µV
0.08%
–
-50 to +500mV
8 µV
0.01%
0.05%
-250 to +250 mV
8 µV
0.01%
0.05%
-1 to +10 V
150 µV
0.01%
0.05%
V
O
LTA
GE
-5 to +5 V
150 µV
0.01%
0.10%
General Conditions – From 15 to 35 degrees C. ambient at the interface board, source resistance less than
1k Ohms. Includes linearity, drift, offset, resolution, and calibration error. 16 bit (0.0015%) resolution in
use.
Thermocouple Conditions – Same as General Conditions. Does not include the accuracy of the
thermocouple itself (cold junction error must be added in; cold junction compensation with T21 terminal
panel only; please see your terminal panel manual’s specifications for cold junction compensation error).
Resolution assumes 16 bit resolution in use; it is approximate because resolution varies with temperature.
Thermocouples use the 50 mV range. For inputs below -5 mV (below approximately -100 degrees C.) use
the +/-25 mV range. For the best resolution use the 50 mV range above -5 mV.