Thermocouple types, Supported product thermocouple ranges, Ranges …… 13-3 – Measurement Computing DataShuttle User Manual
Page 103

DataShuttle and DynaRes
11-12-01
Analog Input & Output 13-3
Select the proper thermocouple for the temperature range and accuracy required using the table below. Use
care in hostile environments especially at very high temperatures. Some thermocouples are durable only in
certain atmospheres. They may fail very quickly in another type of atmosphere. In hostile environments,
the thermocouple is usually placed in a sheath or thermowell.
Thermocouple Types
Alloy Combination
Type
+ Lead
- Lead
E
Nickel-Chromium
(Chromel)
Copper-Nickel
(Constantan)
J
Iron
Copper-Nickel
(Constantan)
K
Nickel-Chromium
(Chromel)
Nickel-Aluminum
(Alumel)
T
Copper
Copper-Nickel
(Constantan)
C
Tungsten – 5% Rhenium
Tungsten – 26% Rhenium
D
Tungsten – 3% Rhenium
Tungsten – 25% Rhenium
G
Tungsten
Tungsten – 26% Rhenium
R
Platinum – 13% Rhodium
Platinum
S
Platinum – 10% Rhodium
Platinum
B
Platinum – 30% Rhodium
Platinum – 6% Rhodium
N
Nickel-Chromium-Silicon
(Nicrosil)
Nickel-Silicon-Magnesium
(Nisil)
Supported Product Thermocouple Ranges
Type
Degrees C
Degrees F
E
-200 to 680
-76 to 1256
J
0 to 750
32 to 1382
K
-200 to 1250
-166 to 2282
T
-200 to 350
-184 to 662
C
0 to 2315
32 to 4200
D
0 to 2315
32 to 4200
G
0 to 2315
32 to 4200
R
0 to 1450
32 to 2642
S
0 to 1450
32 to 2642
B
100 to 1700
212 to 3092
N
-270 to 1300
-450 to 2372