Select the proper thermocouple for the temperature range and accuracy required using the table below. Use
care in hostile environments especially at very high temperatures. Some thermocouples are durable only in
certain atmospheres. They may fail very quickly in another type of atmosphere. In hostile environments,
the thermocouple is usually placed in a sheath or thermowell.

Thermocouple Types

Alloy Combination

Type

+ Lead

- Lead

E

Nickel-Chromium
(Chromel)

Copper-Nickel
(Constantan)

J

Iron

Copper-Nickel
(Constantan)

K

Nickel-Chromium
(Chromel)

Nickel-Aluminum
(Alumel)

T

Copper

Copper-Nickel
(Constantan)

C

Tungsten – 5% Rhenium

Tungsten – 26% Rhenium

D

Tungsten – 3% Rhenium

Tungsten – 25% Rhenium

G

Tungsten

Tungsten – 26% Rhenium

R

Platinum – 13% Rhodium

Platinum

S

Platinum – 10% Rhodium

Platinum

B

Platinum – 30% Rhodium

Platinum – 6% Rhodium

N

Nickel-Chromium-Silicon
(Nicrosil)

Nickel-Silicon-Magnesium
(Nisil)

Supported Product Thermocouple Ranges

Type

Degrees C

Degrees F

E

-200 to 680

-76 to 1256

J

0 to 750

32 to 1382

K

-200 to 1250

-166 to 2282

T

-200 to 350

-184 to 662

C

0 to 2315

32 to 4200

D

0 to 2315

32 to 4200

G

0 to 2315

32 to 4200

R

0 to 1450

32 to 2642

S

0 to 1450

32 to 2642

B

100 to 1700

212 to 3092

N

-270 to 1300

-450 to 2372