beautypg.com

Measurement Computing DataShuttle User Manual

Page 122

background image

Appendix A-2

1

1-13-01

DataShuttle and DynaRes

• You can calibrate to your own temperature measurement by adjusting CCAL in the Calib.Dat file. Add

75 points to the number for every needed 1°C increase in the reading. Starting at 22,000 for CCAL is a
good idea.

• If you see very negative temperatures (like -700 F) then you probably have an open thermocouple. The

ends of the thermocouple must be re-attached, though care must be taken when doing this. If intermit-
tent low temps are encountered then the source is probably noise. On a type K thermocouple 4µV can
cause 0.1°C change. This means that 360mV will show up as ±9,000°C.

• For long thermocouple or extension wire, a shield should be used. This will reduce noise. A shield

consists of a conductive conduit, through which the thermocouple wire is passed. This conduit is in
turn connected to ground (GND not GND(A)). In this way any noise is transmitted through the shield
to ground instead of through the thermocouple wire.

• Thermocouple wire should not be excessively long. Instead of using long wires, attach extension wire

matched to the type of thermocouple. If the extension wire is twisted pair, then noise reduction will be
increased.

This manual is related to the following products: