beautypg.com

Current, Current …… 13-10 – Measurement Computing DataShuttle User Manual

Page 110

background image

13-10 Analog Input & Output

11-12-01

DataShuttle and DynaRes

Figure 13-5. Connections for RTDs

Current

Figure 13-6 shows the connections for current measurement. R3 is used as a shunt resistor. Since these
analog input boards are configured to measure voltage, use the following equation to measure current:

Current = Voltage

÷ R3

Do not run currents higher than 1 Amp into the terminal panel or you risk melting the circuit board traces.
Observe the power rating of the resistor you install at R3. Space is provided for a ¼ Watt resistor. To
measure 1 Amp it can be no larger than 0.25 ohm.

This manual is related to the following products: