beautypg.com

6 hysteresis setting – KEYENCE EX-V Series User Manual

Page 95

background image

Chapter 4 Use of Data Processing Functions

4

85

4.6

Hysteresis Setting

With the EX-V series, the hysteresis for the comparator output (HIGH/GO/LOW
OUT) is factory-adjusted to 5 digits.
Change the hysteresis by following the procedures below when necessary.

1. Press the

key.

The sensor enters the tolerance limit setting mode.

2. Press the

key to display "x".

3. Press the

key.

The sensor enters the hysteresis setting mode.

4. Press the

or

key to set the desired hysteresis value.

5. Press the

key.

The sensor saves the specified hysteresis value and returns to the
measurement mode.

Timing diagram

The hysteresis can be set within the range of 0 to +19999 digits.

SET

CALL

HIGH

TIM

GO

LOW

FUNC

UTILITY

MODE

CALIB

OFF
ON

OFF
ON

OFF
ON

(v)

(t)

HIGH OUT

GO OUT

LOW OUT

HIGH limit value

LOW limit value

Hysteresis