Chapter 2 Quick Guide to Basic Mode Operation
2
27
2.5
Checking the Tolerance Limit Values (Calling the tolerance
limit values)
The preset tolerance limit values can be checked during measurement.
1. Press the
key.
The preset setting number ("0" through "3") is displayed.
Note: Press the
key for a short time. When the key is pressed for 2 seconds or
longer, the sensor enters the adjustment mode. If you accidentally put the sensor
into the adjustment mode, "0" appears at the right end on the display. At this time,
press the
key to return to the measurement state.
2. Press the
key.
The HIGH indicator flashes and the preset HIGH limit value appears on the
display.
3. Press the
key again.
The LOW indicator flashes and the preset LOW limit value appears on the
display.
4. Press the
key to return to measured value display.
Note 1: The EX-V series internally performs measurement and comparison even
while displaying the tolerance limit value. When the measured value exceeds the
tolerance limit value while the tolerance limit value is called out on the display, the
displayed value changes from green to red to indicate that the measured value is
out of the tolerance range.
Note 2: When you want to change the preset value while calling out the tolerance
limit value on the display, press the
key. This allows you to enter the tolerance
limit setting mode. Refer to "2.4 Tolerance Limit Value Setting" (
➮ pages 25 and
26).