Chapter 2 Quick Guide to Basic Mode Operation

2

19

The measurement mode has been factory-set to manual.

➮ For the initial setting of the functions for each measurement mode, refer to "4.1 Table of Data

Processing Functions" (page 46).

4. Press the

key to save the measurement mode setting.

The sensor returns to the measurement state.

Note: Changing the measurement mode initializes all preset values except for the
tolerance limit values.

5. Activate the machine and check the runout value during normal operation.

The value displayed in this step is the runout per cycle.

6. Set the upper value for the tolerance limit.

Calculate the upper limit value by adding the upper limit value of the required
tolerance to the value displayed during normal operation, and set the value as
the HIGH limit value.

Example

When the runout value during normal operation is "10

µm" and the tolerance is

"+30

µm", set the HIGH limit value to "0.040". When the runout value exceeds "40

µm", the comparator output is turned on.

1) Press the

key.

• The sensor enters the tolerance limit setting mode.
• The setting number is displayed.

Note: Press the

key for a short time. When the key is pressed for 2 seconds or

longer, the sensor enters the function setting mode and displays "Ab(d". If you
accidentally put the sensor into the function setting mode, press the

key again

to return to the measurement state.

2) Press the

key to display the desired setting number.

• Pressing the

key changes the setting number sequentially.

• "x" indicates the hysteresis setting (

➮ page 85).

SET

CALL

HIGH

TIM

GO

LOW

FUNC

UTILITY

MODE

CALIB

Setting No.

key