KEYENCE EX-V Series User Manual
Page 32
Chapter 2 Quick Guide to Basic Mode Operation
2
22
4. Press the
key to save the measurement mode setting.
The sensor returns to the measurement state.
Note: Changing the measurement mode initializes all preset values except for the
tolerance limit values. If you accidentally put the sensor into the function setting
mode, press the
key again to return to the measurement mode.
5. Set the upper value for the tolerance limit.
•
Place the reference target in the correct measuring position and check the
normal thickness measurement.
•
Calculate the upper limit value by adding the upper limit value of the required
tolerance to the normal thickness measurement, and set the value as the
HIGH limit value.
Example
When the normal thickness measurement is "1 mm" and the tolerance is "
±0.1
mm", set the HIGH limit value to "1.100" and the LOW limit value to "0.900".
1) Press the
key.
• The sensor enters the tolerance limit setting mode.
• The setting number is displayed.
Note: Press the
key for a short time. When the key is pressed for 2 seconds or
longer, the sensor enters the function setting mode and displays "Ab(d". If you
accidentally put the sensor into the function setting mode, press the
key again
to return to the measurement state.
2) Press the
key to display the desired setting number.
• Pressing the
key changes the setting number sequentially.
• "x" indicates the hysteresis setting (
➮ page 85).
Reference: Four kinds of tolerance limit values ("0" to "3") can be saved in the
memory. The registered setting numbers can be switched using external signals (
➮
page 4).
3) Press the
key.
• The sensor automatically enters the HIGH limit value setting mode.
• The setting value is displayed in red and the HIGH indicator illuminates.
SET
CALL
HIGH
TIM
GO
LOW
FUNC
UTILITY
MODE
CALIB
Setting No.
key
HIGH
TIM
GO
LOW