beautypg.com

KEYENCE EX-V Series User Manual

Page 20

background image

Chapter 1 Preparation

10

1

3. Bring the sensor head in contact with the target.

4. Press the

key.

The sensor displays "2" and moves to the next setting step.

5. Move the sensor head.

Move the sensor head so that the distance between the sensor head and target
is the full scale of the measuring distance (maximum distance) of the sensor
head.

6. Press the

key.

The sensor displays "3" and moves to the next setting step.

7. Move the sensor head.

Move the sensor head so that the distance between the sensor head and target
is half the measuring distance (half scale).

8. Press the

key.

The sensor displays "End".

9. Press the

key to complete the adjustment.

The sensor starts measurement.

Reference: To stop the adjustment, press the

key. Then, the sensor starts

measurement without making the adjustment.

Note: If the distance between the sensor head and target is improperly adjusted
during the adjustment, "Err 6" is displayed. Press the

key to cancel the error,

and then readjust the distance.

SET

CALL

Full scale

SET

CALL

Half scale

SET

CALL