KEYENCE EX-V Series User Manual
Page 33

Chapter 2 Quick Guide to Basic Mode Operation
2
23
4) Press the
or
key to display the desired value.
•
Pressing the
key changes the flashing digit to the right.
•
Pressing the
key changes the value.
•
All digits flash when the fifth digit and sign can be changed.
5) Press the
key.
The HIGH limit setting value is saved. The sensor automatically enters the
LOW limit value setting mode.
6. Set the lower value for the tolerance limit.
•
Set the lower limit value.
•
Use procedures 3) and 4) in step 5 to set the LOW limit value. Then, press
the
key to save the value.
Note 1: The output is disabled during the tolerance limit value setting. When the
output is already on, however, the output is retained.
Note 2: Pressing the
key does not save the changes if [HIGH setting value -
Hysteresis] is smaller than [LOW setting value + Hysteresis]. In this case, "Err1" is
displayed. Press the
key again to cancel the error, and then specify the correct
value.
Reference: When the thickness/gap mode is selected, in the initial setting state the
measured value becomes larger as the distance between the sensor head and
target is shorter. When you want to make the measured value smaller as the
distance is shorter, set function "d" (inverse/normal display function) to normal.
For details, see page 73.
HIGH
TIM
GO
LOW
No setting
(5th digit)
(1st to 4th digits)
HIGH
TIM
GO
LOW
Direction in which the
measured value increase