Chapter 4 Use of Data Processing Functions

4

70

Sampling period
Specify the length of the sampling period after the hold-reset input.

The sampling delay time and sampling period are specified with time length.
The unit of the setting display is "ms".

Example

1000.1 = 1000.1 ms

Flicker time

Turns ON the internal trigger signals in a specified cycle. Set the flicker time to
repeat the measurement continuously in a specific cycle.

The flicker time is specified with time length.
The unit of the setting display is "ms".

Note 1: The sampling delay and sampling period can be specified within the range
of 0.0 to 19999.9 ms.

Note 2: The flicker time can be specified within the range of 0.1 to 19999.9 ms.

Note 3: When "b-0" is selected as the measurement type, only "[-0" is displayed
to be set.

Setting function "["

1. Press the

key for at least 2 seconds.

The sensor enters the common function selection mode.

2. Press the

key to make function "[" flash.

3. Press the

or

key to select the function setting mode.

4. Press the

key to select the desired function.

Select one of settings "0" to "3" and follow the procedures according to the
selected function number.

Sampling period

Trigger point

Starting point

The measured value is
displayed/output at this point.

(Time)

Sampling
delay time