Metrohm 844 UV/VIS Compact IC User Manual
Page 97

4.2 Maintenance and servicing
844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051
85
through the upper opening in the valve housing. The
individual components are rinsed with dist. water and/or
acetone, and the sapphire sphere cleaned with a paper
towel. The valve is then reassembled in accordance with
Figure 28. The components of the inlet and outlet valves are
identical, they are distinguished only by the positioning of the
ceramic holder 151 (see Figure 28).
• Valves that fail to function faultlessly after such cleaning must
be replaced.
• In the reinstallation of the inlet valve 141 or the outlet valve
142 on no account must the two outwardly identical valves be
interchanged. To determine which valve is which, note that
the liquid flows through the pump head from the bottom up.
The flow direction of the valves can be checked simply by
blowing through the clean valve. Both valves are installed with
the black face in the direction of the pump head (see Figure
28).
If by mistake an inlet valve
is installed instead of the outlet valve
, an extreme pressure buildup occurs within the working cylinder,
which is not detected by the pressure transducer and will destroy the
piston seal
!
6 Mounting the pump head
• Reassemble the components of the pump head 27/54 as
shown in Figure 28. Tighten the screw
133 by hand. First
screw in piston cartridge 137 manually until the stop is
reached and then use a wrench to turn it through a further
30°. Firmly tighten the two valve screw holders
143 with a
wrench.
• Reattach pump head 27/54 to the pump with the help of the
4 fixing screws. Firmly tighten them with the 6.2621.030 Allen
key.
To ensure that the pump head is not positioned wrongly, the
holes at the rear for the clamping bolts have different depths,
i.e. 1 clamping bolt is longer than the rest. The deepest hole
must naturally accommodate the longest bolt. If this is not the
case, the pump will not function properly.
• Screw connection capillary 30 to pump head 27/54.
• Connect aspirating tubing 82 to aspirating capillary 30 at the