2.11 Connection of external devices

844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051

57

Net» Software Instructions for Use, Section 6.26.3.

7 Creating a time program for the 844

• Open the window with the system settings of the 844 (with a

double-click on the 844 icon or via the right-hand mouse key

open

).

• Open the

Program

tab.

• Create a time program:

The Remote command sends a signal to the 838 via

Remote line 3. This signal tells the 838 (which is waiting after the

Scan command) to continue with its command sequence.

The two Valve commands switch the injection valve at the

844 to the positions Fill and Inject.

For more detailed information about the control please refer

to Section 3.3.1 or the «IC Net» Software Instructions for Use.

2.11.2 Connection of the 813 Compact Autosampler

The 813 Compact Autosampler, which is optionally available from Me-
trohm, is an automatic sampler for ion chromatography. The 813 Com-

pact Autosampler (Remote connection) is connected to Remote
connection 23 of the 844 UV/VIS Compact IC using the 25-pin
6.2141.130 Remote cable. The control of the 844 UV/VIS Compact IC
takes place using Remote commands in the time program of the 844

UV/VIS Compact IC.
Proceed as follows:

1 Electrical connection 844 – 813

Connect Remote connection 23 of the 844 UV/VIS Compact

IC and the Remote connection of the 813 Compact Auto-

sampler using the 6.2141.130 Cable (see 813 Instructions for

Use).

2 Tubing connection 813 – Injection valve

• Screw off aspirating tubing 41 from injection valve 46/60.
• Cut the PEEK capillary tubing mounted on the 813 Compact

Autosampler to the required length (so that it can be connec-

ted to injection valve

46/60. of the 844).

• Pull the free end of the PEEK capillary tubing through one of