Maintenance and servicing, General information, Shutdown – Metrohm 844 UV/VIS Compact IC User Manual

Page 92: Changing separating columns, 2 maintenance and servicing

background image

Table of contents

4.2 Maintenance

and

servicing

4.2.1 General

information

Care
The 844 UV/VIS Compact IC requires proper care and attention.

Excessive contamination of the instrument could possibly lead to mal-

functions and a shorter service life of the inherently rugged mechanical

and electronic parts.
Spilled chemicals and solvents should be wiped up immediately. It is

especially important to protect the plug connections at the rear of the

instrument (particular the mains plug) against contamination.

Although constructional measures have been designed to virtually

eliminate such a situation, should corrosive media penetrate the

interior of the instruments the mains plug of the 844 UV/VIS Compact

IC must be immediately disconnected to prevent extensive damage to

the instrument electronics. Inform Metrohm service if your

instrument(s) have been damaged in such a way.

The instrument must not be opened by untrained personnel. Please

comply with the safety notes in Section 1.4.1.

Maintenance by Metrohm service
Maintenance of the 844 UV/VIS Compact IC is best done as part of an

annual service performed by specialists from the Metrohm company. If

work is frequently performed with caustic and corrosive chemicals, it

may be necessary to shorten the interval between servicing.
The Metrohm service department is always willing to offer expert advice

on the maintenance and servicing of all Metrohm instruments.

4.2.2 Shutdown

If the 844 UV/VIS Compact IC is shut down for a considerable length of

time, the entire IC system (without column) must be rinsed free from

salt with methanol/water (1:4) to avoid crystallization of eluent salts with

the corresponding subsequent damage.
The connections to the column must be removed for rinsing; the two

inlet and outlet capillaries are connected directly with a coupling.
Before taking it back into operation, rinse the system with eluent

(before the connection of the separation column).

4.2.3

Changing separating columns

It is recommended that the precolumn is also changed each time you

change the separation column). In principle the precolumn requires

changing more frequently then the separating column.

Identical separation system