3.2.4 Instrument
icon
When you click with the right mouse button on the instrument icon, you
will see the following menu:
Open
Opens the window for
System Settings.
Hardware
Opens the window for
Hardware Settings.
Spectrum window
Opens the
SPECTRUM WINDOW
.
Unlink
Unlink device from the system.
3.2.5
Hardware start/stop and baseline recording
Start
Starting the hardware (by selecting
SYSTEM / Control / Startup
hardware
) on the 844 UV/VIS Compact IC includes starting the high-
pressure pump and, depending on the system start values
Startup with
hardware
defined on the
Configuration
tab, starting the UV lamp, the VIS
lamp, and, if present, the column heating and peristaltic pump.
At the same time Recording the measuring signal is started, using
the method defined in the system. The measuring signal is recorded
independently from the time for the chromatogram set under
Duration
,
until either data acquisition is stopped with
SYSTEM / Control / Stop data
acquisition
or a new determination is started. Alternatively data recording
can also be interrupted by clicking on in the chromatogram
window. In this case the user is asked whether the recorded baseline is
to be saved or not.
Stop
By selecting
SYSTEM / Control / Shutdown hardware
the high-pressure
pump, UV lamp, VIS lamp and, if present, the column heating and the
peristaltic pump of the 844 UV/VIS Compact IC will be stopped
immediately. The current determination and the processing of the active
sample table will also be stopped. In addition, the remote output lines
will be set to the values defined for the parameter
PowerON values
on
the tab
Outputs
in the
HARDWARE
window.
3.2.6 Determination
start/stop
Start
By selecting the menu item
SYSTEM
/
Control
/
Start determination
a
determination is started with the settings of the selected system. At the
start the system start values
Startup with hardware
defined for the system