3.3 Settings

844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051

71

Event

Events for automatic remote signal output:

Inject

Switching of the injection valve to the "INJECT"

position.

Fill

Switching of the injection valve to the "FILL"

position.

Pump stop (P max/min)

Pump stopped because pressure limits are

exceeded.

Hardware error

Hardware error detected at the 844 UV/VIS

Compact IC (high-pressure pump, injection

valve, or peristaltic pump not working correctly).

Alarm leak detector

Internal

leak.

PowerOn values

Output

lines

1… 8 will be set to the values

defined here after the instrument has been

switched on or after an emergency stop with

Shutdown hardware

.

Selection:

0

,

1

Pulse length

Length of a pulse in ms.

Alarm stops

The

Alarm stops

tab of the hardware settings window defines the events

for which the instrument is stopped immediately. At an alarm stop, high-
pressure pump and peristaltic pump are stopped immediately, the
running determination and the active sample queue are also stopped.