Table of contents

Malfunction

Cause

Rectification

Extreme peak

broadening,
splitting (double
peaks)

• Dead volume at the co-

lumn ends

• Check the connections

(use PEEK capillaries with
0.25 mm i.d. between
injection valve and
detector block)

• Turn separation column (if

allowed), or replace it

No feed of re-
agent for the
post column
reactor

• Contact pressure too

low

• Leak in system

• Defective pump tubing

• Contamination of the

filter in the 6.2821.120
Filter unit PEEK

• Adjust contact pressure

(see Section 2.10.2)

• Check connections and

make leakproof

• Replace pump tubing (see

Section 4.2.5)

• Replace the 6.2821.130

Filter (see Section 2.3.5)

Bleeping of the
gas sensor

• Gas in the instrument

• Turn off instrument
• If solvent leaked into the

device, remove it.