Metrohm 844 UV/VIS Compact IC User Manual
Page 93

4.2 Maintenance and servicing
844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051
81
If you wish to replace an IC separating column by a column of the same
type, proceed as follows:
1 Remove old column
• Switch off high-pressure pump and wait for pressure to drop.
• With column heating: screw off knurled screws 116, remove
• Screw off the column connection capillaries (both sides) from
the column or precolumn.
2 Connect new column to injection valve
• Remove end caps from column 113.
• Screw inlet end of column 113 (observe flow direction) onto
column inlet capillary 47/110 or onto the precolumn (see
Section 2.8.1).
3 Rinse column
• Place beaker beneath the column outlet.
• Switch on high-pressure pump and rinse column with eluent
for approx. 10 min, then switch off pump.
4 Connect column to detector block
• Screw outlet end of column 113 onto column outlet capillary
(depends on Version: 122(2.844.0010/2.844.0110), 123
(2.844.0020/2.844.0120) or
124 (2.844.0030/2.844.0130)).
Changing the separation system
If you wish to replace an IC separating column by a column of a diffe-
rent type, proceed as follows:
1 Remove old column
• Switch off high-pressure pump and wait for pressure to drop.
• With column heating: screw off knurled screws 116, remove
• Screw off the column connection capillaries (both sides) from
the column or precolumn.
2 Rinse with eluent
• Place a beaker beneath column inlet capillary 47/110.
• If the new eluent is not miscible with the previous eluent then
you must rinse the IC system for approx. 15 min with a
solvent in which both the eluents are miscible.
• Rinse the IC system for approx. 15 min with the eluent
required for the column that is later to be used (flow rate
1 mL/min).
3 Connect new column to injection valve
• Remove end caps from column 113.
• Screw inlet end of column 113 (observe flow direction) onto
column inlet capillary47/110 or onto the precolumn (see
Section 2.8.1).