2.10 Putting into operation

844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051

53

• Select the

Connect to workplace

item from the

Control

menu of

the system window.

2 Open control window

• Double-click on the 844 icon in the system window. The

control window for manual operation of the 844 UV/VIS

Compact IC appears. Open the subtab

Controls

and set the

system parameters (see Chap 3.3.1).

Make sure that the set flow does not exceed the maximum

permissible flow for the particular column (see column data

sheet).

3 Start system

• Make sure that aspirating tubing 82 for the high-pressure

pump is immersed in the eluent, and that aspirating tubing

108 for the post column reactor is immersed in the reagent.

• In the

Control

menu in the system window select the item

Startup hardware (Measure baseline)

. The high-pressure pump

starts up. If they have been selected on the tab

Configuration

(see Section 3.3.1) then the peristaltic pump and the column

heating will also start up. The chromatogram window of the

set data recorder opens and the baseline(s) is/are shown

continuously.

4 Set contact pressure for pump tubings

• On tubing cartridge 96 press contact pressure lever 97

upward until the reagent solution is aspirated.

• Then press contact pressure lever 97 upward by 1 more click

in order to achieve an optimal contact pressure.