Metrohm 844 UV/VIS Compact IC User Manual
Page 77

3.3 Settings
844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051
65
Pressure max, MPa
Value for maximum switch-off pressure for high-
pressure pump. Once it has been set this value
will be monitored, even if there is no connection
to the PC (units depend on the settings made
under
IC NET /
Global preferences
).
Entry range:
0.0
...
35.0 MPa
If the
Flow
is higher than 1.5 mL/min, the defined
Pressure max
must not exceed 25.0 MPa (250 bar).
Pressure min, MPa
Value for minimum switch-off pressure for high-
pressure pump. Once it has been set this value
will be monitored, even if there is no connection
to the PC (units depend on the settings made
under
IC NET /
Global preferences
).
Entry range:
0.0
...
35.0 MPa
Column oven
Column
oven
Turn column oven on (only enabled, if a column
oven is connected).
Turn column oven off (only enabled, if a column
oven is connected).
Temperature, °C
Set value for temperature in the column oven.
Entry range:
0 ... 75 °C
Remote lines
Set values for remote line settings 1…8.
Selection:
0
,
1
Detector
On the subtab
Detector
the UV/VIS detector 43/61 can be controlled
manually. The parameters for the measurement can also be set:
UV lamp
UV
lamp
Turn UV lamp on.
Turn UV lamp off.