Page 88

background image

Table of contents

refakt

Int

refzero

Int

akt

Int

zero

Int

Abs

λ

λ

λ

λ

log

log

1

1

=

where:

refzero

Int

λ

The intensity measured when

recording the zero spectrum at the
wavelength set for the reference
channel.

refakt

Int

λ

The actual intensity measured at

the wavelength set for the reference
channel.

Intensity

The intensity (y-axis) is given as a function of

the wavelength (x-axis).

Zero

A click on

Zero

shows the last zero spectrum to

have been recorded (a zero spectrum is

recorded by clicking on the

button on

the subtab

Manual/Detector

of the 844 control

window).
The zero spectrum is a reference spectrum. It
should be recorded at a time at which there is
no sample in the measuring cell. It is used for
calculating the absorption spectrum (see above
formulas).

Wavelength, nm

Sets the shown x-axis range

Entry range:

190 ... 900 nm

, (standard settings

190 nm and 900 nm,

Upper limit

must be higher

than

Lower limit

).

Spectrum

Sets the shown y-axis range

Standard settings:

0 ... 100.

Spline

Activates curve smoothing.

Send values to the 844 UV/VIS Compact IC.

Opens the dialog for entering the path and file

name. Saves the last recorded spectrum in the

selected .cvs-file. Format: 2 lines of floating

numbers (first line: wavelength, second line:

absorption or intensity).

Copies plot into clipboard.

Closes

window.

If a right-hand mouse click is made in the spectrum window the
wavelength

Current wavelength

and the corresponding absorption value

Current spec. value

will be shown.