Page 47

2.7 Post column reactor
844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051
35
2 Attach coupling 93
• At the left-hand end of pump tubing 94 6.1826.320 push on a
coupling 93 6.2744.030. Slightly moistening the Tipp of the
coupling makes this easier.
• Fix "Aspirating tubing for reagent" 92 to coupling 93 using
compression fitting
64 (6.2744.010).
3 Attach "Coupling with filter and tubing security
• To the right-hand end of pump tubing 94 6.1826.320
attach a "coupling with filter and tubing security device" 99
6.2744.180:
• Dismantle the tubing security device and first push the
union nut and compression piece onto the tubing.
• Push the tubing onto the coupling and screw the union
nut onto the coupling to secure the tubing.
4 Insert the pump tubing
• Insert the fitted-out pump tubing 94 into the tubing cartridge.
The middle stopper 95 should click into position in the
corresponding holder on left-hand side of the tubing
cartridge.
5 Reinsert the tubing cartridge
• Insert the tubing cartridge with pump tubing 94 into the front
of the holding clip. Insert the left-hand side first, then press
down the right-hand side until snap-action lever 98 clicks into
position. Make sure that the pump tubing has no kinks.
Figure 17: Installing pump tubings
64 Compression fitting (6.2744.010)
72 Filter 6.2821.130
92 Aspirating tubing for reagent
6.1803.020 PTFE tubing
93 Coupling (6.2744.030)
96
97
64 93
92
94 95
98
95
64
99
100
72