3 spectrum window, Spectrum window – Metrohm 844 UV/VIS Compact IC User Manual

Page 86

background image

Table of contents

T int, ms

Integration

time.

Entry range:

10

...

254 ms

T out, s

Measuring time per data point.

Entry range:

0.05

...

2.00 s

Data rate

Effective measuring time per data point. The

Data rate

is the particular whole number multiple

of the value

T int

that is closest to the value

T

out

.

The measurement of a data point is explained by using an

example:
Set values:

T int

= 15 ms,

T out

= 0.1 s

The calculated

Data rate

is 105 ms (the particular multiple of

T

int

that is closest to the value

T out

; in our example 7 times 15

ms). This means that the integration will be carried out 7 times

(each time for 15 ms) and then averaged to give a data point.

Optimizes

T int

and adjusts the VIS lamp.

It should be done

• after exchanging a lamp (UV or VIS)
• if parts of the intensity spectrum are in the

overload ( > 60'000 , peaks cut)

• if the maximum of the intensity spectrum

of dist. water is too low (below 33'000)

Before clicking

, the flow-through

cell should be rinsed and filled with dist.

water. There must be no air bubbles in the

flow-trough cell when clicking

.

Operating hours

Operating hours of the lamps.

UV lamp, h

Display of the operating hours of the
UV lamp.

Set operating hours of the UV lamp
to 0. Click it after replacing the UV
lamp.

VIS lamp, h

Display of the operating hours of the
VIS lamp.

Set operating hours of the VIS
lamp
to 0. Click it after replacing the

VIS lamp.

3.3.3 Spectrum

window

The

SPECTRUM WINDOW

can be opened with a right-hand mouse-

click on the 844 icon and then selecting the menu item

Spectrum

window

, or via the

button on the subtab

Manual/Detector

.