Metrohm 844 UV/VIS Compact IC User Manual

Page 80

background image

Table of contents

List of program instructions

The following program instructions can be added to the time program
on the

Program

subtab:

Remote

0, 1, *, p

Set remote output lines 1...8 to the
desired values. For entry of the first
value, enter

1

,

0

,

p

or

*

. For entry of

the other values, move the cursor in
front of the value to be changed and
enter

1

,

0

,

p

or

*

.

Valve Inject,

Fill

Switch injection valve to "INJECT"
or "FILL" position.

ICPump on,

off

Switch on or off the high-pressure
pump
.

Flow

0.2 ... 2.5 mL/min

Set flow rate of the high-pressure
pump to the desired value.

Pmax

0.0 ... 35.0 MPa

Set maximum pressure limit for the
high-pressure pump to the desired

value.

Pmin

0.0 ... 35.0 MPa

Set minimum pressure limit for the

high-pressure pump to the desired
value.

Peristaltic on,

off

Switch on or off the peristaltic
pump
.

Wavelength_1..3

190 ... 900 nm

Set wavelength of the correspon-
ding measuring channel on the
desired value.

Bunching_1..3

1 ... 101 nm

Set bunching of the corresponding
measuring channel on the desired
value.

UV_Lamp on,

off

Switch on or off the UV lamp.

VIS_Lamp on,

off

Switch on or off the VIS lamp.

Zero

Record

a

new

zero spectrum.

Remote configuration

On the

Remote configuration

subtab user-defined remote commands can

be defined, which can be inserted into a time program.