Metrohm 844 UV/VIS Compact IC User Manual
Page 80

Table of contents
List of program instructions
The following program instructions can be added to the time program
on the
Program
subtab:
Remote
0, 1, *, p
Set remote output lines 1...8 to the
desired values. For entry of the first
value, enter
1
,
0
,
p
or
*
. For entry of
the other values, move the cursor in
front of the value to be changed and
enter
1
,
0
,
p
or
*
.
Valve Inject,
Fill
Switch injection valve to "INJECT"
or "FILL" position.
ICPump on,
off
Switch on or off the high-pressure
pump.
Flow
0.2 ... 2.5 mL/min
Set flow rate of the high-pressure
pump to the desired value.
Pmax
0.0 ... 35.0 MPa
Set maximum pressure limit for the
high-pressure pump to the desired
value.
Pmin
0.0 ... 35.0 MPa
Set minimum pressure limit for the
high-pressure pump to the desired
value.
Peristaltic on,
off
Switch on or off the peristaltic
pump.
Wavelength_1..3
190 ... 900 nm
Set wavelength of the correspon-
ding measuring channel on the
desired value.
Bunching_1..3
1 ... 101 nm
Set bunching of the corresponding
measuring channel on the desired
value.
UV_Lamp on,
off
Switch on or off the UV lamp.
VIS_Lamp on,
off
Switch on or off the VIS lamp.
Zero
Record
a
new
zero spectrum.
Remote configuration
On the
Remote configuration
subtab user-defined remote commands can
be defined, which can be inserted into a time program.