Metrohm 844 UV/VIS Compact IC User Manual

Page 70

background image

Table of contents

the openings of the 844 UV/VIS Compact IC and use a

6.2744.010 PEEK Compression fitting to screw it onto

injection valve 46/60..

3 Tubing connection Injection valve – Waste

• Loosen connection capillary 33 from injection valve 46/60.

(sample outlet).

• Cut a piece of 6.1831.060 PTFE capillary (sample changer

accessory) to the required length, attach it to injection valve

46/60. using a 6.2744.010 PEEK compression fitting and

place the other end in a waste container.

4 Setting a method on the 813

• Use the 813 keyboard on the 813 Compact Autosampler to

set the required method (see 813 Instructions for Use).

5 Creating a time program for the 844

• In the main window of the «IC Net» software select

File / Open /

System

. Open the system for the 844 UV/VIS Compact IC that

you have created in Section 2.5.4..

• Open the window with the system settings of the 844 UV/VIS

Compact IC (with a double-click on the 844 icon or via the

right-hand mouse key

open

).

• Open the

Program

tab.

• Create a time program:

The Remote command sends a signal to the 813 via

Remote line 3. The function of this signal depends on the

method selected at the 813 (see 813 Instructions for Use,

Section 3.4).

The two Valve commands switch the injection valve of the

844 to the positions Fill and Inject.

For more detailed information about the control please refer

to Section 3.3.1 or the «IC Net» Software Instructions for Use.