Metrohm 844 UV/VIS Compact IC User Manual
Page 78

Table of contents
VIS lamp
VIS
lamp
Turn VIS lamp on.
Turn VIS lamp off.
Records a new zero spectrum.
The zero spectrum is a reference spectrum. It
should be recorded at a time at which there is
no sample in the measuring cell. It is used for
calculating the absorption spectrum.
Opens
the
SPECTRUM
window (see Section
Wavelength, nm
Sets the wavelengths at which the
chromatograms are to be recorded (for max.
three measuring channels).
Bunching, nm
Sets the bunching of the wavelengths defined
above.
Channel 1, nm
Measuring
channel
1.
Channel 2, nm
Measuring
channel
2.
Channel 3, nm
Measuring
channel
3.
The relationship between
Wavelength
and
Bunching
is explained by using
an example:
Set values:
Wavelength
= 500 nm,
Bunching
= 21 nm
An upper and a lower limit for the measured wavelengths of the
corresponding measuring channel are calculated from these values.
For these calculations the following equation applies: subtract "1"
from the
Bunching
value, divide the new value by 2, and subtract/add
the resulting value from/to the
Wavelength
value.
Result: wavelengths between 490 nm (lower limit) and 510 nm
(upper limit) will be included in the chromatogram.
Time program
On the
Program
tab of the system settings window a user-defined time
program for instrument control can be entered. This program is started
automatically as defined in the
Start mode
window (see Section 4.3.3)
either at the moment the determination is started (
Start with
determination
) or at the moment the sample is injected (
Start with inject
).