5 appendix, Technical data, 1 uv/vis detector – Metrohm 844 UV/VIS Compact IC User Manual
Page 104: Appendix, Uv/vis detector, Sect. 5, 1 technical data

Table of contents
5 Appendix
5.1 Technical
data
Unless otherwise specified, the published data
comprises typical values for the 844 UV/VIS Compact IC
at an ambient temperature of 25 °C.
5.1.1 UV/VIS
Detector
Measuring range
Absorption
-2.0 ... 2.0 AU
Number of channels
3 absorption channels + 1 ratio channel
Resolution
2.5 * 10-7 AU
Noise
2.5 * 10-5 AU (data rate 1/s))
Drift
at reference conditions
5 * 10-4 AU/h
Wavelength range
Wavelength range
λ
190...900 nm
Bunching
Δλ
1...limit nm (limit is 2 times the difference bet-
ween the wavelength of the measuring channel
and the nearer wavelength border (190 resp. 900
nm).
Δλ
is always odd.)
Accuracy absolute
+/- 3 nm
Stability
+/- 1 nm
Optical resolution
5 nm
Measuring interval
Data rate for each channel 0.5 ... 20 samples/sec
Analog output
Voltage range
-1.0 ... 1.0 V
Resolution
30 μV
Noise
<30 μV
Lamps
UV lamp
Type
Power
consumption
Lifetime
D
2
(Deuterium)
approx. 20 ... 30 W
approx. 1000 h
VIS lamp
Type
Power
consumption
Lifetime
Halogen
approx. 5 W
approx. 10'000 h