Metrohm tiBase 1.1 User Manual
Page 34
2.4 Formula editor
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
tiBase 1.1
Variable
name
(PC-Control)
Variable
name
( tiBase 1.1)
Description
Command
.FD#
.FP{x}.TIM
Time for fixed endpoint x (1…9)
in s
DET, MET, SET, KFT, KFC,
STAT, MEAS (without T/
Flow)
.FP#
.FP{x}.[VOL]
Volume for the fixed endpoint x
(1...9) in mL
DET, MET, SET, KFT, STAT
.HM#
.HP{x}.MEA
Measured value for HNP x (1...9)
in mV
1)
DET, MET
.HT#
.HP{x}.TEM
Temperature for HNP x (1...9) in
°C
DET, MET
.HD#
.HP{x}.TIM
Time for HNP x (1...9) in s
DET, MET
.HP#
.HP{x}.[VOL]
Volume for HNP (half neutraliza-
tion potential) x (1...9) in mL
DET, MET
#M.MIM
.IME
Initial measured value (measured
value before start conditions are
processed) in the unit of the
measured value (number)
DET, MET, SET, KFT, STAT,
MEAS, CAL MEAS,
STDADD, DOS
.MIT
.ITE
Initial temperature (temperature
before start conditions are pro-
cessed) in °C
DET, MET, SET, KFT, STAT,
MEAS (without T/Flow),
CAL MEAS, STDADD, DOS
.XAM
.MA.MEA
Maximum measured value in the
unit of the measured value
DET, MET, SET, KFT, STAT,
MEAS
.XAT
.MA.TEM
Temperature for the maximum
measured value in °C
DET, MET, SET, KFT, STAT,
MEAS
.XAD
.MA.TIM
Time for the maximum measured
value in s
DET, MET, SET, KFT, STAT,
MEAS
.XAP
.MA.[VOL]
Volume with maximum meas-
ured value in mL
DET, MET, SET, KFT, STAT
.MI.GFL
Minimum gas flow rate in
mL/min
MEAS T/Flow
.XIM
.MI.MEA
Minimum measured value in the
unit of the measured value
DET, MET, SET, KFT, STAT,
MEAS
.XIT
.MI.TEM
Temperature for minimum meas-
ured value in °C
DET, MET, SET, KFT, STAT,
MEAS
.XID
.MI.TIM
Time for the minimum measured
value in s
DET, MET, SET, KFT, STAT,
MEAS
.XIP
.MI.[VOL]
Volume for minimum measured
value in mL
DET, MET, SET, KFT, STAT
.RMC
.MR.MRC
Correlation coefficient for mean
dosing rate for the whole range
STAT, DOS