beautypg.com

13 determinations - displaying history, 14 determinations - make current, 15 determinations - show calibration curve – Metrohm tiBase 1.1 User Manual

Page 261: Displaying determination history

background image

■■■■■■■■■■■■■■■■■■■■■■

4 Database

tiBase 1.1

■■■■■■■■

253

4.5.2.13

Determinations - Displaying history

Menu item: Database

Determinations Show history

Switch history view on/off
With the menu item Determinations

Display history... or the symbol

only the currently focused determination in the Determination table as

well as all the previous versions of this determination will be shown.

If the History view with the menu item Determinations

Display his-

tory... or the symbol

is disabled again, then the original selection of

determinations in the Determination table will appear again.

4.5.2.14

Determinations - Make current

Menu item: Database

Determinations Make current

Make old version current
With the menu item Determinations

Make current... or the symbol

the determination version selected in the determination table will again

be made the current determination version. This creates a new determina-
tion whose version number is increased by +1 compared with the last ver-
sion to have been saved.

4.5.2.15

Determinations - Show calibration curve

Dialog window: Database

Determinations Show calibration curve… Calibra-

tion curve

With the Determinations, Show calibration curve… menu item or
the symbol the calibration or standard addition curve is shown for the
determination selected in the Calibration curve dialog window.

Calibration curve for CAL LOOP pH
In the Calibration curve dialog window a tab marked with the com-
mand name showing the calibration curve and calibration data will be dis-
played for each CAL LOOP pH command . The command type is shown
above the calibration curve. The curve display shows the measured values
and the curve calculated from these measured values. Listed below are the
individual measured values and the results for Slope, E(0) and Variance.

Calibration curve for CAL LOOP Conc
In the Calibration curve dialog window a tab marked with the com-
mand name showing the calibration curve and calibration data will be dis-
played for each CAL LOOP Conc command . above the calibration curve.
The curve display shows the measured values and the curve calculated
from these measured values. Listed below are the individual measured val-
ues and the results for Slope, E(0), c(blank) and Variance. Calibration
curve for CAL LOOP Conc