beautypg.com

3 curve evaluation - endpoint list – Metrohm tiBase 1.1 User Manual

Page 244

background image

4.5 Determination overview

■■■■■■■■■■■■■■■■■■■■■■

236

■■■■■■■■

tiBase 1.1

Define properties for the graph display

The dialog window Properties - Curves # is opened for graph
display. The curve properties defined for each command type are
saved for the reprocessing window per client.

Show EP list

The dialog window Endpoints is opened in which all endpoints
of the selected curve are displayed in a table. If an endpoint is
added to or deleted from the curve then the table will be updated
automatically.

Show coordinates
The current coordinates of the cursor are shown in the coordinate display:

X:

X-coordinate.

Y:

Y-coordinate.

4.5.2.6.4.6.3

Curve evaluation - Endpoint list

Dialog window: Database

Determinations Reprocess... Reprocessing Mod-

ifications

Curve evaluation [Edit] Curve evaluation Show EP list End-

points

The dialog window Endpoints opened with the icon

shows all auto-

matically found and manually set endpoints of the selected curve in tabu-
lar form. The endpoints are sorted according to ascending volume, if an
endpoint is deleted from or added to the curve then the table will be
updated automatically.

Endpoint

Designation of the endpoint with number. The following names are possi-
ble:

Selection

EP#

EP#
Endpoint that has been determined automatically by potentiometric
evaluation or set manually by using the intersection or tangent
method.

Selection

BP#

BP#
Endpoint that has been determined by automatic break point evalua-
tion.

Selection

FP#