2 General program functions
tiBase 1.1
■■■■■■■■
25
Variable
name
(PC-Control)
Variable
name
( tiBase 1.1)
Description
Command
.MEN
.ENP
Electrode zero point for the sen-
sor used for the command
(dimensionless for pH sensor or
in mV for the ISE sensor, num-
ber) and/or electrode zero point
calculated from the calibration
(for STDADD and CAL LOOP)
DET pH, DET U, MET pH,
MET U, SET pH, SET U,
STAT, MEAS pH, MEAS U,
MEAS T, MEAS Conc, CAL
LOOP, DOS, STDADD
.EP{x}.DME
Measured value differential for
the endpoint x (1…9)
MET
.EF#
.EP{x}.ERC
ERC for the endpoint x (1…9)
DET
.EM#
.EP{x}.MEA
Measured value for the endpoint
x (1…9) in the unit of the meas-
ured value
DET, MET, SET, KFT, KFC
.ESI#
.EP{x}.MEP
Marking for the endpoint x
(1...9); 1 = 1 endpoint, 2 = 2 or
more endpoints
DET, MET, SET, KFT
.EP#
.EP.QTY
Measured value (water) for the
endpoint in mg
KFC
.ET#
.EP{x}.TEM
Temperature for the endpoint x
(1...9) in °C
DET, MET, SET, KFT
.ED#
.EP{x}.TIM
Time for the endpoint x (1...9) in
s
DET, MET, SET, KFT, KFC
.EP#
.EP{x}.VOL
Volume for the endpoint x (1…
9) in mL
DET, MET, SET, KFT
.MCT
.ETE
End temperature (temperature
following processing of the com-
mand) in °C
DET, MET, SET, KFT, KFC,
STAT, MEAS (without T/
Flow), CAL MEAS,
STDADD, DOS
.MCV
.EVT
End volume (total dosed volume
at the end of the command) in
mL
DET, MET, SET, KFT, STAT,
DOS
.FM#
.FP{x}.MEA
Measured value for the fixed
endpoint x (1…9) in the unit of
the measured value
DET, MET, SET, KFT, STAT
.FP{x}.QTY
Measured value (water) for the
fixed endpoint x (1...9) in mg
KFC
.FT#
.FP{x}.TEM
Temperature for the fixed end-
point x (1…9) in °C
DET, MET, SET, KFT, STAT,
MEAS (without T/Flow)