2.4 Formula editor

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

tiBase 1.1

Command variables
Unless marked otherwise, all of the variables listed here in alphabetical
order are of the Number type.

Note

In the case of variables with Index {x}, the desired number 1…9 must
be entered for x (e.g. .EP{3}.ERC for the third endpoint).

If no index is specified, then the last index will be used automatically
(e.g. .EP.ERC for the last endpoint).

Variable
name

(PC-Control)

Variable
name

( tiBase 1.1)

Description

Command

.BM#

.BP{x}.MEA

Measured value for the break
point x (1…9) in the unit of the
measured value

DET, MET, MEAS (without
T/Flow)

.BT#

.BP{x}.TEM

Temperature for the break point
x (1…9) in °C

DET, MET, MEAS (without
T/Flow)

.BD#

.BP{x}.TIM

Time for the break point x (1…9)
in s

DET, MET, MEAS (without
T/Flow)

.BP#

.BP{x}.VOL

Volume for the break point x
(1…9) in mL

DE, MET

.CONC

.CONC

Concentration of the solution
used for the command (number)

DET, MET, SET, KFT, STAT,
STDADD, ADD, DOS, LQH

.MCD

.DBL

Duration command (total dura-
tion for the processing of the
command) in s

DET, MET, SET, KFT, KFC,
STAT, MEAS, CAL MEAS,
STDADD, DOS

.MDC

.DRI

Current and/or last drift for drift
correction in mL/min

SET, KFT, KFC

.MSD

.DSC

Duration start conditions (time
for processing all start condi-
tions) in s

DET, MET, SET, KFT, STAT

.DDC

.DTI

Time for drift correction (time
from the start of the titration to
the end of the command) in s

SET, KFT, KFC

.MCM

.EME

End measured value (measured
value after processing of the
command) in the unit of the
measured value

DET, MET, SET, KFT, KFC,
STAT, MEAS, CAL MEAS,
STDADD, DOS