Bump Testing a Gas Tracer • 94

If at any point during the bump test the flow rate to the instrument becomes
too low, the PC Controller Program will abort the bump test and display a
failed bump test screen. The BUMP

T

LED on the SDM-2012 control

panel will be solid red. Click OK to return to the main program window or
double click an instrument information line to view the bump test or
calibration results in the Logs window. The BUMP

T

LED on the SDM-

2012 control panel will turn off. Confirm that all tubing connections are
correct and that all lines are clear.

Figure 74: Low Flow Bump Test Failure

24. For all bump test result outcomes, the instruments will be displayed in the main

program window in Details View. To return the main program window to Icons
View, click the Icons radio button.

25. All bump test results can be viewed in the Logs window. See “Bump Test Data”

on page 138 for instructions for viewing bump test data.