111 • Calibrating a Gas Tracer

If at any point during the calibration the flow rate to the instrument becomes
too low, the PC Controller Program will abort the calibration and display a
failed calibration screen. The CAL

S

LED on the SDM-2012 control panel

will be solid red. Click OK to return to the main program window or double
click the instrument information line to view the calibration results in the
Logs window. The CAL

S

LED on the SDM-2012 control panel will turn off.

Confirm all tubing connections are correct and that all lines are clear.

Figure 89: Low Flow Calibration Failure

23. For both calibration result outcomes, the instrument(s) will be displayed in the

main program window in Details View. To return to Icons View, click the Icons
radio button.

24. When a calibration is performed, any bump test or calibration data that is being

stored in the instrument will be downloaded to the PC Controller Program. This
data can be viewed in the Logs window along with the results of the calibration
just performed. See “Bump Test Data” on page 138 or “Calibration Data” on
page 125 for instructions for viewing bump test or calibration data.