background image

91 • Bump Testing a Gas Tracer

23. After the fresh air purge is complete,

If the bump test passes, the following window will open indicating that. The
BUMP

T

LED will be solid green. Click OK to return to the main program

window. The BUMP

T

LED will turn off.

Figure 71: Bump Test Passed

The Gas Tracer will automatically be shut off 20 seconds after a successful
bump test.