79 • Bump Testing a Gas Tracer

8.

You will be asked to confirm which Gas Tracer channels will use GAS 1 for the
bump test. The gas concentration shown next to each channel is the auto
calibration value for that channel. If it does not match the gas concentration in
the cylinder(s), change the auto calibration value for that channel. See “Sensor
Tab” on page 165 for instructions. If you are bump testing a standard Gas
Tracer, leave the first channel (the ppm sensor) deselected and select all other
channels. If you are bump testing a standard Gas Tracer plus a %volume
sensor, leave the first and last channels (the ppm and %volume sensors)
deselected and select all other channels.

Figure 60: Gas Cylinder Confirmation Screen

Standard Gas Tracer Standard Gas Tracer Plus %Volume