RKI Instruments SDM-2012 PC Controlled Configuration User Manual
Page 190

185 • Bump Testing GX-2012s and Gas Tracers
If you are bump testing a 0 - 10000 ppm Gas Tracer, the screen will appear as
shown below. The ppm sensor will be bump tested using GAS 2. The GAS 2
selection box is deselected by default. You must select the GAS 2 box to bump
test the ppm combustible sensor. If you do not select the GAS 2 box, the bump
test will not continue. Be sure that the ppm calibration cylinder is connected to
the GAS 2 fitting with a 6 inch humidifier at the start of the bump test.
Figure 146: Gas Order Confirmation Screen, 0 - 10000 ppm Gas Tracer
10. Click OK. You will be asked if you want to bump test the instruments. Click OK
to proceed. If you don’t want to proceed with the bump test, click Cancel to
return to the main program window.
- Eagle Series (89 pages)
- Eagle Series (66 pages)
- 01 Series Manual (33 pages)
- 01 Series Quik Start (2 pages)
- OX-94 (13 pages)
- FP-3 (16 pages)
- GasWatch 1 (18 pages)
- GasWatch 2 Manual (37 pages)
- GasWatch 2 Quik Start (2 pages)
- GP-01 (34 pages)
- GP-01 with self resetting alarms (33 pages)
- GP-204 (7 pages)
- NP-204 (13 pages)
- RI-215A (24 pages)
- SC-01 (45 pages)
- SC-01 (68 pages)
- Data Cal 2000 (88 pages)
- EAGLE 2 Manual (309 pages)
- Eagle 2 Quik Start (2 pages)
- GX-94 (3 pages)
- GX-94 (52 pages)
- SDM-E2 (89 pages)
- SDM-E2 (224 pages)
- Eagle 1 Manual (89 pages)
- Eagle 1 Quik Start (2 pages)
- Inert Gas Testing Eagle with Two Pumps (66 pages)
- Gas Tracer (234 pages)
- Gas Tracer 10,000 ppm (120 pages)
- GX-2001 Manual (45 pages)
- GX-2001 Quik Start (2 pages)
- GX-2001 Canadian Version (43 pages)
- GX-2003 Manual (94 pages)
- GX-2003 Quik Start (2 pages)
- GX-2009 Manual (73 pages)
- GX-2009 Quik Start (2 pages)
- GX-2009 Supplement (2 pages)
- GX-2012 Manual (185 pages)
- GX-82 (22 pages)
- GX-82A (43 pages)
- GX-86 (22 pages)
- GX-86A (39 pages)
- GX-8000(PT0-098) (52 pages)
- GX-8000(H4-0050) (23 pages)
- RX-516 (36 pages)
- GX-2009 (2 pages)