Shunt Error Status

Software Utilities

Shunt Error Status

This is the error status for each bridge from the last shunt cal that was performed.
The error status is a bit-mapped status word, where the error is present if the bit is
set to 1.

Bit 0 = Bridge amplifier was saturated prior to applying the shunt.

Bit 1 = Bridge amplifier was saturated after applying the shunt.

Bit 2 = Measured delta value failed high.

Bit 3 = Measured delta value failed low.

Bit 4 = A procedural error occurred (unable to read a value, etc.).

In the following example the error status received is 5 indicating that bits 0 and 2
were set.

Bit

4

3

2

1

0

Bit Value

16

8

4

2

1

Example

0

0

1

0

1

Error Status=

5