beautypg.com

MTS SWIFT 20 Ultra Sensor User Manual

Page 50

background image

SWIFT 20 Sensors

50

TIXFER - Low-Profile Transducer Interface Transfer

Software Utilities

The bridge and angle zero values will change whenever a zero is activated by
pressing the Low-Profile TI front panel zero button. Therefore, after a zero is
performed the zero values uploaded will not match those downloaded.

All parameters within the Low-Profile TI have been calibrated. Because the
resolution of each adjustable parameter is finite, downloaded values may vary
slightly from subsequent uploads and the values uploaded from various units may
be different. The downloaded value gets “snapped” to the closest realizable
value. The realizable values depend on the resolution and calibration data for the
parameter. Normally the gains from the SWIFT automated calibration stand are
computed to a higher resolution than the Low-Profile TI is capable of dealing
with.