IP

N 07

4-

28

9L

Composer Operating Manual

1.7 Electrical Power Requirements

The following sections detail the electrical power requirements for the
Composer as minimally required by, or supplied by the optional Power Module.

WARNING

Failure to comply with the electrical power
requirements stated below may result in the unit
malfunctioning or being damaged, and could result in
personal bodily injury.

1.7.1 Rated Output to Composer

Voltage . . . . . . . . . . . . . . . . . . . . . 24 V(dc) ±5%, isolated from ground

Ripple and Noise . . . . . . . . . . . . . 1 V peak to peak max.

Overvoltage Transient Protection . 36 V

Current . . . . . . . . . . . . . . . . . . . . . 2 A min. continuous at 50 °C ambient

Turn on Surge . . . . . . . . . . . . . . . . 35 A

Operational Peak Loads . . . . . . . . 2.5 A for 10 seconds max.

Short Circuit Protection . . . . . . . . . Required, VFB, VIFB, PWM allowed

Short Circuit Time . . . . . . . . . . . . . Indefinite

Ambient Operating Temperature . . 0 to 50 °C

Humidity . . . . . . . . . . . . . . . . . . . . 0 to 95% non condensing

24 V Connector Interface . . . . . . . Custom Shielded Cable Assembly

Connector . . . . . . . . . . . . . . . . 9 Pin D-Sub-Miniature Female

Length . . . . . . . . . . . . . . . . . . . 5 ft. ± 1ft. (1.5 m ± 0.3 m)

Wiring Size . . . . . . . . . . . . . . . #20 AWG

+24 V . . . . . . . . . . . . . . . . . Pins 6, 7, 8

+24 V ret . . . . . . . . . . . . . . Pins 1, 2, 3

Shield Drain . . . . . . . . . . . . Pin 9

Shell . . . . . . . . . . . . . . . . . Isolated