1 system and hardware errors, 2 system and hardware warnings – INFICON Composer Gas Concentration Controller User Manual
Page 153

7 - 11
IP
N 07
4-
28
9L
Composer Operating Manual
7.4.4.2.1 System and Hardware Errors
0x8000 . . . . . . . . . . . . . . . . . . . . . Non Volatile Memory failed
0x4000 . . . . . . . . . . . . . . . . . . . . . TI failed to initialize
0x2000 . . . . . . . . . . . . . . . . . . . . . Temperature chip failed to initialize
0x1000 . . . . . . . . . . . . . . . . . . . . . DSP Initialization failure
0x0800 . . . . . . . . . . . . . . . . . . . . . Temperature chip lockup
0x0400 . . . . . . . . . . . . . . . . . . . . . Heater failure/disconnect
0x0200 . . . . . . . . . . . . . . . . . . . . . Temperature out of range
0x0100 . . . . . . . . . . . . . . . . . . . . . Temperature A to D failure
0x0080 . . . . . . . . . . . . . . . . . . . . . No signal
0x0040 . . . . . . . . . . . . . . . . . . . . . Concentration result error
0x0020 . . . . . . . . . . . . . . . . . . . . . A good fit could not be found
0x0010 . . . . . . . . . . . . . . . . . . . . . Carrier and Precursor gases chosen are
identical
7.4.4.2.2 System and Hardware Warnings
0x0100 . . . . . . . . . . . . . . . . . . . . . Temperature A to D failure
0x0040 . . . . . . . . . . . . . . . . . . . . . Concentration result error
0x0008 . . . . . . . . . . . . . . . . . . . . . Bad peak shape - indicates possible
problem with readings
0x0004 . . . . . . . . . . . . . . . . . . . . . Gas combination could result in ambiguous
results
0x0002 . . . . . . . . . . . . . . . . . . . . . DSP took too long to respond
NOTE: Warnings 0x100 and 0x0040 are counterparts to the same numbered
errors. These conditions are not serious unless the condition persists.
If they persist, an error will be generated.
Likewise, with warnings 0x0008 and 0x0002, if the condition generating
the condition is serious and persistent, other errors — most likely error
0x0080, 0x0040, or 0x0020 — will be generated.
Warning 0x0004 will generate an Error 0x0040 if it enters a region of
the curve when the combination is ambiguous.