background image

4 - 27

IP

N 07

4-

28

9L

Composer Operating Manual

4.3.6.7 RS-232C

See

Chapter 7

for information regarding RS-232C.

4.3.6.8 CONCENTRATION OUT Connector

A BNC female connector that provides a precision analog voltage from 0 to
10.000 V into a 2 K

load that is proportional to either the concentration or

frequency shift. It operates with 15 bit equivalent resolution, monotonic to 1 LSB

4

CONCENTRATION INput
function

0 = ignore,
1 = control to target concentration equals

where:
input voltage range is set by SW-7
max = value of parameter

MAX INPUT (USER#5)

min = value of parameter

MIN INPUT (USER#6)

diff = (max-min)

5

Control Dilution Flow

0 = ignore
1 = maintain Dilution Flow as (Total Flow
- Bubbler Flow) any time the controller is
on and the Control or Bypass command
has been given, unless Input #3 (STOP)
of the system I/O is active. Used in
Concentration Control configuration only.

6

spare

7

Input Voltage Range

0 = 0-10 volts
1 = 0-5 volts
Selects input voltage range for both
CONTROL INput BNC (see

section

4.3.6.10 on page 4-28

) and the optional

Analog I/O accessory’s input; see

section

Table 3-3 on page 3-13

for Bulk Flow or

section Table 3-5 on page 3-19

for

Concentration.

8

Control Mode

0 = Concentration Control configuration
(standard)
1 = Bulk Flow Control configuration;
requires optional Analog I/O card.

Table 4-5 Configuration SW#2 Descriptions (continued)

SW

Function

Details

min input diff

V

in

input voltage range

----------------------------------------------

+