INFICON Composer Gas Concentration Controller User Manual

Page 117

background image

4 - 21

IP

N 07

4-

28

9L

Composer Operating Manual

S. . . . . . . . . . . . . . . . . . . . . . . . . . Pressing this key will temporarily convert

the numerical display’s function to showing
the measured resonant frequency’s
amplitude (millivolt RMS @ 1 V RMS
drive).

T . . . . . . . . . . . . . . . . . . . . . . . . . . Pressing this key will temporarily convert

the numerical display’s function to showing
the resonant chamber’s temperature (°C).

Q . . . . . . . . . . . . . . . . . . . . . . . . . Pressing this key will temporarily convert

the numerical display’s function to showing
the measured resonant frequency’s quality
factor.

F . . . . . . . . . . . . . . . . . . . . . . . . . . Pressing this key will temporarily convert

the numerical display’s function to showing
the current measured resonant frequency
(Hz).

D . . . . . . . . . . . . . . . . . . . . . . . . . Concentration Control configuration:

pressing this key will temporarily convert
the numerical display’s function to showing
the current dilution flow (sccm).

Bulk Flow Control configuration:
temporarily displays bulk flow.

B . . . . . . . . . . . . . . . . . . . . . . . . . Pressing this key will temporarily convert

the numerical display’s function to showing
the bubbler flow readback.

5 . . . . . . . . . . . . . . . . . . . . . . . . . . Pressing this key will temporarily convert

the numerical display’s function to show
the current warnings. (Only active if the
WARNING indicator is illuminated.)

8 . . . . . . . . . . . . . . . . . . . . . . . . . . Concentration Control configuration:

pressing this key temporarily displays the
current target Concentration as interpreted
from the applied voltage on the CONTROL
IN connector.

Bulk Flow Control configuration:
pressing this key temporarily displays
the target Bulk Flow as interpreted from
the applied voltage on the CONTROL IN
connector.