beautypg.com

1 index, 2 bubbler mfc (sccm) – INFICON Composer Gas Concentration Controller User Manual

Page 169

background image

8 - 15

IP

N 07

4-

28

9L

Composer Operating Manual

8.3.5.1 Index

Program Number. . . . . . . . . . . . . The Composer instrument can store up to

nine sets of parameters under this index
field. Thus, it is important that this field be
appropriately selected before downloading
any parameter. This field can be altered
only when the control state is in Stop
position. When this field is modified (1
through 9), the Get button is automatically
pressed to retrieve the chosen parameter
set, which becomes the current parameter
set.

8.3.5.2 Bubbler MFC (sccm)

Full Scale: . . . . . . . . . . . . . . . . . . A numeric data field representing the

maximum flow range (sccm) of the Mass
Flow Controller used to control bubbler’s
gas flow. This parameter accepts values
from 0 to 5000.

Max: . . . . . . . . . . . . . . . . . . . . . . . A numeric data field representing the

maximum flow to be allowed. This
parameter accepts numbers from 0 to 2000
(sccm) and must not exceed Full Scale
value.

Min: . . . . . . . . . . . . . . . . . . . . . . . A numeric data field representing the

minimum flow to be maintained. This
parameter accepts numbers from 0 to
2000.0 (sccm) and must not exceed Max
value.