Composer Operating Manual

The eight inputs are active when the specific input’s terminal is pulled to ground
(common) (<0.8 V) through a contact closure or with TTL/CMOS logic having
current sink capability of 2 ma. (1 low power TTL load). The input ports are
protected to +/- 30 V with respect to chassis ground. See

Table 4-2

.

NOTE: These ports are updated every 200 milliseconds; signals must be

present during a read cycle. All inputs are leading edge detected and
executed.

.

Table 4-2 System I/O Connector Function

Relay
I/O #

FUNCTION

PIN #s

Comment

K1

@ TEMPERATURE

1,2

Closed when the Resonant
Chamber is within +0.1

°C to

-0.2

°C of set point

K2

ERROR

3,4

Opens when the instrument
is in an error. See

Table 6-7

on page 6-7

.

K3

@ CONCENTRATION

5,6

Closed when concentration
is within programmed target
band

K4

STEADY

7,8

Concentration Control
configuration:
closed when
front panel STEADY light is
illuminated.
Bulk Flow Control mode:
used to pass control voltage
from Control Out of
Composer to bubbler’s MFC
when in external control.

K5

user set Reference Zero

9,10 NO
10,11 NC

Concentration Control
configuration:
closed when
factory Reference Zero has
been overridden
Bulk Flow Control mode:
used to pass control voltage
from reactor to bubbler’s
MFC when in external
control. Use pins 10,11 to
insure continuity if instrument
is powered off.