4 category 4; installation related – INFICON Composer Gas Concentration Controller User Manual

Page 136

background image

6 - 4

IP

N 07

4-

28

9L

Composer Operating Manual

6.2.4 Category 4; Installation Related

Table 6-4 Category 4; Installation Related

Failure

Symptom Visibility

Effect on Accuracy

Acoustically noisy
environment

Noisy C signal

Noise on C signal

Electrically noisy
environment

Noisy C signal

Noise on C signal

Maximum flow rate
exceeded

Noisy C signal

Noise on C signal

Inlet and outlet reversed

None

None

Ambient temperature
exceeds setpoint

Temperature and
Warning LEDs blink,
Error relay opens

N/A

Low line voltage

HALT lamp illuminates

N/A

Analog output's load
<2K ohm

None

Nonlinear response of
control and C outputs

Noise on or variability of
the analog input

None, (control hunting
not visible without
Monitor software for
viewing)

Proportional to input
variation